FCC Class 15 Part B
TEL7001 carbon dioxide detector
Technical specifications
Patent method for detection of double beam infraredabsorption method
Sampling diffusion sampling or circulation sampling (50-100ml/min)
CO2 channel
The measurement range of 0-10000ppm display, 0-4000ppm output voltage
The sensitivity of 1ppm
The accuracy of ± 50ppm and ± 5%
Repeated + 20ppm
Temperature coefficient of ± 0.1% readings or ± 2ppm/ ℃ (reference temperature of 25 ℃)
Pressure coefficient /mmHg (0.13% readings can enter the altitude adjustment)
Years of drift ± 20ppm
The response time of T90<60s (22 ℃)
Preheating time <60s (22 ℃)
Working conditions of 0-50 ℃, 0-95%RH, without dew
The storage temperature of -40-60 ℃
The calibration interval of 12 months, offset adjustmentusing 0-1000ppm CO2
Temperature channel
Display the temperature range of 0-50 ℃, 0-40 ℃ voltage output
The display resolution of 0.1 ℃ (0.1 ℉)
Display options C, f or closed, panel button settings
The accuracy of ± 1 ℃ (2 ℉)
The response time of 20-30min (depends on the ambient temperature)
The calibration interval of 12 months, offset adjustment using standard temperature (10-30 ℃)
Analog output
CO2 0-4VDC, 1mV/ppm (4000ppm)
Temperature of the linear 0-4VDC, 0-40 ℃ (32-104 ° f)
The output impedance of 100 ohms
Connect cable RJ-45 to DB9 adapter cable
Display LCD alone showed CO2/ temperature readings (panel button to set the altitude, C / F, calibration function)
Power Supply
Battery type 4 AA batteries
Battery life: 80 hours (alkaline)
External power supply 6VDC external AC/DC adapter
Power requirements 100mA peak, the average value of 20mA, 6V
Certification FCC Class 15 Part B